Projection Choice and Organ Dose

Organs on the beam's entrance side get the most dose. A PA projection puts the breasts, thyroid and eyes on the low-dose exit side.

Entrance vs exit

The x-ray beam weakens as it crosses the body, because tissue absorbs and scatters it. The entrance skin dose, where the beam enters, is the highest dose in the field. Through an adult abdomen or lumbar spine, only about 1 percent or less of the entrance dose reaches the exit surface, so the entrance dose can be 100 times the exit dose or more. Even the air-filled chest lets only a small fraction through. An organ near the entrance surface therefore absorbs far more dose than the same organ near the exit surface, even though both are inside the field.